Product Details
Product Details
Product Specification
Usage |
1. Sample processing and requirements: 4、EnzymeKnotCombinethingWorkWorkLiquidmatchManufacture: Centrifuge 100 × concentrated enzyme conjugate at 1000 × g for 1 minute 15 minutes before use, dilute 100 × concentrated HRP enzyme conjugate to 1 × working concentration with universal diluent (example: 10 uL concentrated solution + 990 uL universal diluent), and prepare for ready use. 5、1×Wash liquid preparation: Take 10mL of 20 × washing liquid into 190mL distilled water (the concentrated washing liquid taken out of the refrigerator may have crystals, which is a normal phenomenon. It can be left at room temperature and prepared after the crystals are completely dissolved). 5. Operation steps: 1. Take out the required slats from the aluminum foil bag after equilibration at room temperature for 10 minutes, and seal the remaining slats with a ziplock bag and put them back to 4 °C. 2. Add samples: Add samples or standards of different concentrations to the corresponding wells at 100uL per well, and add 100uL of universal diluent to the blank wells. Incubate at 37 °C for 60 min after covering the plate sealing film. (suggest : Will waitThe test sample is diluted at least 1 times with universal diluent and then added to the enzyme labeled plate for testing.So as to reduce the influence of matrix effect on the test results, and finally, the sample concentration needs to be multiplied by the corresponding dilution factor when calculating. It is recommended to set up double wells for all samples and standards to be tested during testing). 3. Add biotinylated antibody: take out the enzyme labeled plate, discard the liquid, and do not wash it. 100 uL of biotinylated antibody working solution was directly added to each well, and the plate sealing membrane was covered and incubated at 37 °C for 60 minutes. 4. Plate washing: Discard the liquid, add 300uL 1x washing liquid to each hole, let it stand for 1 minute, throw away the washing liquid, pat dry on absorbent paper, and repeat washing the plate 3 times (you can also use a plate washing machine to wash the plate). 5. Add enzyme conjugate working solution: Add 100uL of enzyme conjugate working solution to each well, cover the sealing membrane and incubate at 37 °C for 30 minutes. 6. Wash the plate: Discard the liquid and wash the plate 5 times according to the washing method in step 4. 7. Add substrate: Add 90uL of substrate (TMB) to each well, cover with a plate sealing film, and incubate at 37 °C in the dark for 15 minutes. 8. Add stop solution: Take out the enzyme plate, directly add 50uL of stop solution to each well, and immediately measure the OD value of each well at a wavelength of 450nm. VI. Calculation of experimental results: Result judgment: 1. Calculate the average OD value of the standard product and the sample double well and subtract the OD value of the blank well as the correction value. Taking concentration as abscissa and OD value as ordinate, the standard curve of four-parameter logic function is drawn on double logarithmic coordinate paper. 2. If the OD value of the sample is higher than the upper limit of the standard curve, it should be properly diluted and retested and multiplied by the corresponding dilution factor when calculating the sample concentration. Typical data and reference curves: The following data and curves are for reference only, and experimenters need to establish standard curves according to their own experiments.
attention : This picture is for reference only The sample content should be calculated from the standard curve drawn from each experimental data. 7. Kit performance: 1. Repeatability: The intra-plate coefficient of variation is less than 10%, and the inter-plate coefficient of variation is less than 10%. 2. Recovery rate: Add 3 different concentration levels of human LZM to the serum and plasma of selected healthy people to calculate the recovery rate.
3. Linear dilution: High concentration human LZM was added to the selected 4 healthy human serum and plasma, and diluted within the kinetic range of the standard curve to evaluate the linearity.
8. Problem analysis: |
|||||||||||||||||||||||||||||||||||||||||||||||||||||
Problem Description | Possible cause | Corresponding countermeasures | ||||||||||||||||||||||||||||||||||||||||||||||||||||
Poor scaling curve | Incorrect dilution of standard | Ensure that the standard is dissolved and diluted according to the recommended method | ||||||||||||||||||||||||||||||||||||||||||||||||||||
Inaccurate pipetting | Calibrate the pipette periodically and check tip tightness | |||||||||||||||||||||||||||||||||||||||||||||||||||||
Evaporation of the reaction solution | Enzyme-labeled plate is sealed with a sealing membrane | |||||||||||||||||||||||||||||||||||||||||||||||||||||
Incomplete plate washing | Sufficient number of washes and addition of sufficient amount of washing liquid | |||||||||||||||||||||||||||||||||||||||||||||||||||||
Foreign matter at the bottom of the hole | Clean bottom of plate before reading | |||||||||||||||||||||||||||||||||||||||||||||||||||||
Weak or colorless chromogenic | Incubation time is not enough | Ensure incubation time | ||||||||||||||||||||||||||||||||||||||||||||||||||||
Incorrect incubation temperature | Incubate at recommended temperature | |||||||||||||||||||||||||||||||||||||||||||||||||||||
Insufficient reagent volume addition | Inspect the pipette and follow the procedure exactly | |||||||||||||||||||||||||||||||||||||||||||||||||||||
Incorrect dilution | Test Reagent Dilution Step | |||||||||||||||||||||||||||||||||||||||||||||||||||||
Enzyme conjugate inactivation | Mixed enzyme conjugate and substrate, checked by color reaction | |||||||||||||||||||||||||||||||||||||||||||||||||||||
Low OD value | Incorrect plate reader settings | Check instrument wavelength | ||||||||||||||||||||||||||||||||||||||||||||||||||||
No stop solution added | Add an appropriate amount of stop solution | |||||||||||||||||||||||||||||||||||||||||||||||||||||
Wait time too long when reading the board | Timely plate reading | |||||||||||||||||||||||||||||||||||||||||||||||||||||
Excessive sample content | The appropriate dilution factor was determined by pre-experiment | |||||||||||||||||||||||||||||||||||||||||||||||||||||
Sample content is too low | The appropriate dilution factor was determined by pre-experiment | |||||||||||||||||||||||||||||||||||||||||||||||||||||
Background height | Contamination of chromogenic solution | Change the color developing solution | ||||||||||||||||||||||||||||||||||||||||||||||||||||
Color development time is too long | Controlling color development time | |||||||||||||||||||||||||||||||||||||||||||||||||||||
Wrong dilution of detection antibody or enzyme conjugate | Use recommended dilution method | |||||||||||||||||||||||||||||||||||||||||||||||||||||
Incomplete plate washing | Sufficient number of washes and addition of sufficient amount of washing liquid |
Name | 9 6T match set | Storage conditions after opening |
Pre-coated 96-well plate | 8 holes × 12 strips | -20℃ |
Standard | 2 sticks | -20 ℃, use on the day after reconstitution |
Universal diluent |
2 × 20 mL |
2-8℃ |
Concentrated biotinylated detection antibody (100 ×) |
120uL |
-20℃ |
Concentrated enzyme conjugate (100 ×) |
120uL |
-20 ℃ (protected from light) |
20 × washing solution |
2 × 10 mL |
2-8℃ |
Substrate (TMB) |
10mL |
2-8 ℃ (protected from light) |
Stop liquid |
6mL |
2-8℃ |
Sealing film |
4 sheets |
without |
Instructions |
1 serving |
without |
2. Incorrect plate washing may lead to inaccurate results. Make sure to drain the liquid from the wells as much as possible before adding the substrate. Do not allow the micropores to dry for too long throughout the process.
3. Clean the residual liquid and fingerprints at the bottom of the plate, otherwise it will affect the OD value.
4. The substrate color development solution should be colorless, and the substrate solution that has turned blue cannot be used.
5. Avoid cross-contamination of reagents and samples to avoid wrong results.
6. Avoid direct exposure to strong light during storage and incubation.
7. Any reaction reagent cannot come into contact with the bleaching solvent or the strong gas emitted by the bleaching solvent. Any bleaching component will destroy the biological activity of the reagents in the kit.
8. Expired products cannot be used, and components with different item numbers and batch numbers cannot be mixed.
9. Recombinant proteins from sources other than the kit may not match the antibodies in this kit and are not recognized.
10. If the disease may be spread, all samples should be managed well, and the samples and testing devices should be handled according to the prescribed procedures.
