Product Details
Product Details
Product Specification
Usage | 1. Preparation of cell culture medium: (1) Take 1 mL of cell culture medium from a cell culture in the logarithmic growth phase. (2) Centrifuge the cell culture medium at 200 g for 5 minutes at room temperature. Transfer approximately 800 μL of the culture medium to a new centrifuge tube, taking care not to carry over any cell debris. (3) Use the prepared cell culture medium as a template for nested PCR for mycoplasma detection. | |
Volume (μL) | ||
2xPCR Mix | 25 | |
3xPCR Mix | 4 | |
5 | ||
6 | ||
7 | ||
8 | ||
9 | 2 | |
Template (cell culture medium) 1 | ||
PCR reaction internal control 2 | Add sterile ultrapure water to | 3 |
4 | ||
5 | ||
6 | ||
7 | ||
8 | ||
9 | ||
1 | ||
PCR reaction conditions: 30 cycles, conditions are shown in the table below:
Step | Temperature | Time | Number of cycles |
Pre-denaturation | center;">94℃ | 5min | 1 |
Degeneration | 94℃ | 30s | 30 |
Annealing | 55℃ | 30s | 30 |
Extension | 72℃ | 40s | 30 |
Final extension | 5min | 1 |
(3) Prepare 2% agarose gel. Take 10 μL of the second PCR product for DNA electrophoresis.
3. Results The internal control of the PCR reaction is a single DNA band of 515 bp. Common mycoplasma contamination will produce a single DNA band of 236 bp-365 bp. Acholeplasma laidlawii contamination will produce double DNA bands of 426 bp and 219 bp.
Product composition:
MYCO-TTM Nested PCR Mycoplasma Detection Kit contains 2 tubes of ready-to-use primers and 1 tube of reaction internal control. The specifications are as follows:
Component | 25T | 50T |
Primer P1 | 50uL | 100uL |
Primer P2 | 50uL | 100uL |
PCR reaction internal control | 50uL | 100uL |
2. PCR reactions should be performed in a clean bench to avoid cross-contamination and aerosol contamination. Nuclease-free pipette tips and reaction tubes should be used, and pipette tips with filters are recommended.
3. Hot-start Taq enzyme series is recommended for PCR reagents. The 2X PCR premix should be free of DNA gel electrophoresis stain.
4. PCR reactions should be performed in a clean bench to avoid cross-contamination and aerosol contamination. Nuclease-free pipette tips and reaction tubes should be used, and pipette tips with filters are recommended.