Store at 2-8°C, protected from light; product shelf life is 12 months.

Background

Homogeneous Immuno Chemiluminescence Assay (HICA) is a homogeneous immunoassay method based on energy transfer between donor beads and acceptor beads at close proximity, resulting in luminescence.

word导入的商品图片

 

Components

Specification

Fill Volume

250 μg

50 μL

5 mg

1 mL

25 mg

1 mL x 5


Protocol

```html

【Required Reagents】

Name

Catalog No.

Anti-Human IgG Fc Acceptor Beads UA086097
Streptavidin Donor Beads UA086104
Universal Buffer 1 UA086113


【Detection Procedure (Reference)】

Procedure

Procedure 1 (37°C Rapid Detection)

Procedure 2 (Room Temperature Detection)

Step 1:

4μL Tag1-M1 +4μL Tag2-M2+ 6μL Acceptor Beads,Light-protected/Green light

4μL Tag1-M1 +4μL Tag2-M2+ 6μL Acceptor Beads,Light-protected/Green light

Incubation

37°C shaking incubation for 20 minutes,Light-protected/Green light

Room temperature incubation for 60 minutes,Light-protected/Green light

Step 2:

Add 6μL Donor Beads,Light-protected/Green light

Add 6μL Donor Beads,Light-protected/Green light

Incubation

37°C shaking incubation for 10 minutes,Light-protected/Green light

Room temperature incubation for 30 minutes,Light-protected/Green light

Reading

Instrument reading

Instrument reading


【Performance Validation】

Sample Preparation:

Biotinylated human IgG (Bio-hIgG) was pre-diluted to 15μg/mL (100nM) using Universal Buffer 1 as the stock solution, followed by serial dilution as below:

ID

Final Concentration (nM)

Universal Buffer 1

Volume (μL)

High Concentration Add

Volume (μL)

C12

1.0E+01

210

90μL stock solution

C11

3.0E+00

210

90μL C12

C10

1.0E+00

180

90μL C11

C9

3.0E-01

210

90μL C10

C8

1.0E-01

180

90μL C9

C7

3.0E-02

210

90μL C8

C6

1.0E-02

180

90μL C7

C5

3.0E-03

210

90μL C6

C4

1.0E-03

180

90μL C5

C3

3.0E-04

210

90μL C4

C2

1.0E-04

180

90μL C3

C1

0

180

/


Detection Reagent Preparation:

Name

Preparation Concentration

Diluent

Anti-Human IgG Fc Acceptor Beads

25 μg/mL

Universal Buffer 1

Streptavidin Donor Beads

25 μg/mL

Universal Buffer 1


37°C Incubation Mode Results:

Max signal: 4772579

Min signal: 856

  EC50= 0.171 nM

Room Temperature Incubation Mode Results:

Max signal: 1912242

Min signal: 375

EC50= 0.155 nM

```

Guidelines

1. This experiment is light-sensitive; ensure light protection during operation. It is recommended to perform preparation, sample loading, and incubation steps under green light (illuminance below 100 LUX). 2. This product is compatible with multifunctional microplate readers equipped with an Alpha detection module. 3. Vortex thoroughly before use or briefly centrifuge (2000×g, 5–10 seconds) to ensure complete usage. 4. It is recommended to use the company's accompanying dilution buffer for reagent preparation and sample dilution. If additional components are required, they can be directly added to this buffer. 5. To ensure comparability of experimental data across different batches, strictly control incubation temperature and time. 6. Avoid generating bubbles during sample loading.