Skip to product information
1 of 1

UniOne® TR-FRET Human KRAS[WT] CYPA Ternary Binding Kit

UniOne® TR-FRET Human KRAS[WT] CYPA Ternary Binding Kit

Catalog Number: UA086156 Brand: UA BIOSCIENCE
Price:
Regular price $730 USD
Regular price Sale price $730 USD
Size:

For shipping services or bulk orders, you may request a quotation.
Secure checkout with
View full details

Product Details

Product Specification


Host Human
Stability & Storage

-80℃

Background

ASSAY PRINCIPLE

The kit is based on homogeneous time-resolved fluorescence technology (TR-FRET) technology,which is used to assess the molecular-mediated interaction between the KRAS[WT] and CypA. This method provides a high-throughput approach to detect the small molecules capable of mediating the interaction between the KRAS[WT] and CypA simply and rapidly.

As shown in the figure below, the interaction between KRAS[WT] and CypA was detected by an Eu-labeled anti-Tag1 antibody (TR-FRET donor) and an Ac-labeled anti-Tag2 antibody (TR-FRET receptor). Since the molecular adhesive Daraxonrasib mediates the interaction between KRAS[WT] and CypA, the donor antibody is in close proximity to the receptor antibody. Excitation of the donor antibody triggers fluorescence resonance energy transfer (FRET) to the receptor antibody, resulting in specific emission of a signal at a wavelength of 665 nm from the receptor antibody. This specific signal is proportional to the degree of interaction between Daraxonrasib, KRAS[WT] and CypA. This homogeneous assay is simple to perform and requires no washing steps.

Components

【Composition and Storage Conditions】

Composition

concentration

100T

500T

2500T

10000T

Storage

Tag1-KRAS[WT] protein

5

8μL

40μL

200μL

800μL

-80℃

Tag2-CypA protein

50×

8μL

40μL

200μL

800μL

-80℃

Daraxonrasib

400 μM

5μL

10μL

50μL

200μL

-80℃

Anti-Tag1 Eu antibody

10

5μL

25μL

75μL

500μL

-80℃

Anti-Tag2 Ac antibody

200×

5μL

13μL

38μL

250μL

-80℃

Detection buffer

10×

400μL

2mL

10mL

40mL

-80℃


Note: Aliquot into appropriate volumes, store at the recommended temperature and avoid repeated freeze-thaw cycle.


【Process and Operations】

1、Reagent Preparation

1.1 Compositions equilibration

Before use, dissolve all reagents at room temperature (allow at least 30 minutes for temperature equilibration). For the 384-well plate assay, use 20 μL of reaction mixture (the reagent quantities are listed in the table below). Calculate the required volume prior to preparation accordingly. The following preparation steps are provided for reference only, using a 500-well plate as an example.

Table 1. Reagent Preparation

Composition

Configuration

μL/well

Detection buffer

Add 2 mL of 10× Detection buffer to 18 mL of deionized water, dilute to 1× Detection , and mix thoroughly for later use.

-

Tag1-KRAS[WT] protein

Take 40 μL of the 50× Tag1-KRAS[WT] protein stock solution, add 1.96 mL of 1× Detection buffer to dilute it to 1× concentration, and mix thoroughly for later use.

4

Tag2-CypA protein

Take 40 μL of 50× Tag2-CypA protein, add it to 1.96 mL of 1× Detection buffer to dilute to 1× concentration, and mix thoroughly for later use.

4

Antibody Mix

Take 25 μL of the 100× Anti-Tag1 Eu antibody stock solution, add 2475 μL of 1× Detection buffer to dilute to 2.5 mL, and mix thoroughly; take 13 μL of the 200× Anti-Tag2 Ac antibody stock solution, add 2587 μL of 1× Detection buffer to dilute to 2.6 mL, and mix thoroughly; combine the two solutions in a 1:1 ratio to obtain the Antibody Mix.

10


1.2 Gradient dilution of the sample

Daraxonrasib as an example, the dilution solution is 1× Detection buffer. To minimize interference from matrix effects, it is recommended to use a solution with the same matrix composition as the sample being tested; additionally, the sample concentration should be adjusted according to its actual situation.

Table 2. Daraxonrasib gradient dilution(Adjust according to the actual situation)

Daraxonrasib final concentration(nM)

Daraxonrasib preparation concentration(nM)

Dilution

400.00

4000.00

1μL 400 μM Daraxonrasib +99μL 1× Detection buffer

200.00

2000.00

30μL +30μL 1× Detection buffer

100.00

1000.00

30μL +30μL 1× Detection buffer

50.00

500.00

30μL +30μL 1× Detection buffer

25.00

250.00

30μL +30μL 1× Detection buffer

12.50

125.00

30μL +30μL 1× Detection buffer

6.25

62.50

30μL +30μL 1× Detection buffer

3.13

31.25

30μL +30μL 1× Detection buffer

1.56

15.63

30μL +30μL 1× Detection buffer

0.78

7.81

30μL +30μL 1× Detection buffer

0.39

3.91

30μL +30μL 1× Detection buffer

Blank

0

0

40μL 1× Detection buffer


2、Sample and control

2.1 Sample:Add 4 μL of Tag1-KRAS[WT] protein working solution, 4 μL of Tag2-CypA protein working solution, 2 μL of the gradient-diluted sample , and 10 μL of the mixed Antibody Mix sequentially into the 384-well plate.

2.2 Daraxonrasib: 4 μL of Tag1-KRAS[WT] protein working solution, 4 μL of Tag2-CypA protein working solution, 2 μL of gradient-diluted Daraxonrasib, and 10 μL of the mixed Antibody Mix sequentially into the 384-well plate.

2.3 Blank:4 μL of Tag1-KRAS[WT] protein working solution, 4 μL of Tag2-CypA protein working solution, 2 μL of Detection buffer, and 10 μL of the mixed Antibody Mix sequentially into the 384-well plate.

2.4 Negative control 10μL Detection buffer, and 10 μL of the mixed Antibody Mix sequentially into the 384-well plate.

After adding all samples, centrifuge, seal with a cover membrane, and incubate at room temperature for 1 hours.

Table 3. Sample and control spotting process

Daraxonrasib

Sample

Blank

Negative control (NC)

2μL

gradient-diluted Daraxonrasib

2μL

gradient-diluted sample

2μL

1× Detection buffer

10μL 1× Detection buffer

and 10μL Antibody Mix

4μL Tag1-KRAS[WT] protein

4μL Tag2-CypA protein

10μL Antibody Mix

Seal the plate wells with a sealing film and incubate at room temperature for 1hours.


3、Read Data

The detection was performed on a TR-FRET microplate reader. The excitation wavelength was 320/340 nm, and the emission wavelengths at 620 nm and 665 nm were detected.

【DATA REDUCTION 】

1) Ratio

Ratio = (665/620) ×10000

2) Net signal

Net signal = (Std-NC)/NC×100

Std:the ratio of sample and control

NC: the ratio of negative control

3) CV(%)

CV(%)= Standard Deviation/Mean Ratio × 100%

Data Example

The following data cannot replace experimental results obtained in experiments; they are provided solely for illustrative purposes, and outcomes may vary depending on the plate reader used.

/3Note: Recommended microplate

(384-well plate, white, shallow wells).