Product Details

Product Specification

protein IL-2
Usage Own test equipment required
1. Plate reader (measurable 450nm The absorption value of the detection wavelength and 540nm or 570nm (Absorption value of corrected wavelength)
2. High-precision filler and disposable suction tip
3. Distilled water or deionized water
4. Bottle washer (spray bottle), multi-channel plate washer or automatic plate washer
5. 500mL measuring cylinder

I. Pre-experimental preparation
1. Sample Collection and Storage
① Cell culture supernatant: particulate matter should be removed by centrifugation; Test the sample immediately. If the sample is not tested in time after collection, it is recommended to repackage it according to the single-use amount and freeze it. -20℃ Inside the refrigerator, avoid repeated freezing and thawing. Samples may require diluent ( 1× Dilute.
② Serum: Use a serum separation tube ( SST Collect samples and let them sit at room temperature. 30 minutes. centrifuge 15 Minutes, rotation speed 1000g 。 Serum was removed immediately and tested immediately. If the sample is not tested in time after collection, it is recommended to repackage it according to the single-use amount and freeze it. ≤-20℃ Inside the refrigerator, avoid repeated freezing and thawing. Samples may require diluent ( 1× Dilute.
③ Plasma: Use EDTA Heparin or citric acid are used as anticoagulants to collect plasma. After collection... 30 Centrifuge within minutes 15 Minutes, rotation speed 1000g And test immediately. If the sample is not tested in time after collection, it is recommended to repackage it according to the single-use amount and freeze it. ≤-20℃ Inside the refrigerator, avoid repeated freezing and thawing. Samples may require diluent ( 1× Dilute.
2. Reagent preparation ( Please allow all reagents and samples to stand at room temperature before use. 15 minutes. All experimental samples and standards are recommended. Perform multiple hole testing )
①1× Washing solution preparation: The concentrated washing solution in the kit is 20× The mother liquor must be diluted with distilled water before use. 1× Working fluid. Example: take 10mL concentrated washing liquid +190mL Distilled water is set to 200mL In actual operation, the usage can be calculated first, and then the preparation can be carried out.
②1× Dilution buffer preparation: The concentrated dilution buffer in the kit is 10× The mother liquor must be diluted with distilled water before use. 1× Working fluid. Example: take 3mL Buffer for concentration and dilution +27mL Distilled water is set to 30mL 。 In actual operation, the required amount of dilution buffer can be calculated based on the sample dilution factor before preparation.
③ To detect antibodies: Centrifuge the dry powder to the bottom of the tube and use... 110uL Dilution buffer ( 1× Dissolve and let stand at room temperature. 5 Get it in minutes 100× mother liquor; Dilute before use 1× Working fluid. According to the dosage per well 100uL Calculate the required volume. Example: Used 10 Each hole, then take 10uL of 100 The working concentration of the detection antibody was determined using dilution buffer ( 1× ) Set the volume to 1mL , get 1mL of 1× Detection antibody at working concentration.
④SA-HRP : SA-HRP For 40× Mother liquor, dilution buffer must be used before use ( 1× Dilute and prepare 1× Working fluid, the amount required per well is 100uL 。 Example: Used 10 Each hole, then take 25uL of 40× mother liquor +975uL Dilution buffer ( 1× ) Set the volume to 1mL , get 1mL of 1× Detection antibody at working concentration.
⑤ Color developer: per well 100uL Calculate the dosage required for the current test, take out the corresponding volume of color developer, and protect it from light; The removed color developer is for same day use only.
⑥ Standard: Dilution buffer for lyophilized standard ( 1× Resolution, resolution volume 1000uL The resulting concentration is 1000pg/mL Standard mother liquor. Gently shake at least 5 Minutes, it fully dissolves. Add to each dilution tube 300uL Dilution buffer ( 1× )。 Make serial dilutions of the standard mother liquor according to the figure below. Each tube must be thoroughly mixed before pipetting to the next tube. The undiluted standard stock solution can be used as the highest point of the standard curve ( 1000pg/mL ), dilution buffer ( 1× ) can be used as the zero point of the standard curve ( 0pg/mL )。



II. Operating Steps
1. Prepare all required reagents and standards;
2. Remove the microwell plate from the sealed bag that has been equilibrated to room temperature. Put the unused strip back into the aluminum foil bag and reseal it.
3. Add to microplate 300uL Washing liquid, let stand and soak. 30 Seconds, discard the lotion and pat the microplate dry on absorbent paper. Use immediately. Do not let the microplate dry.
4. Add standards, experimental samples, or quality controls of different concentrations to the corresponding wells, respectively. Each well... 100uL 。 Seal the reaction wells with sealing tape and incubate at room temperature. 2 hour ;
5. Suck the liquid out of the plate and wash the plate using a bottle washer, multi-channel plate washer, or automatic plate washer. Add washing liquid to each well 300uL Then, the washing liquid in the plate is sucked away. repetitive operation 3 Next. Removing as much residual liquid as possible during each plate wash will help obtain good experimental results. After the last washing of the plate, please dry all the liquid inside the plate or turn the plate upside down and pat dry any remaining liquid on absorbent paper;
6. Add into each microwell 100uL Antibodies are detected. Seal the reaction wells with sealing tape and incubate at room temperature. 2 hours;
7. Repeat number 5 Step-by-step washing plate operation
8. Add into each microwell 100uLSA-HRP Incubation at room temperature 20 minutes. Pay attention to avoiding light;
9. Repeat number 5 Step-by-step washing plate operation
10. Add into each microwell 100uL Chromogenic solution, incubated at room temperature 5-30 Minutes, take care to protect from light;
11. Add into each microwell 50uL Stop solution, the color of the solution in the well will change from blue to yellow. If the solution turns green or the color changes inconsistently, gently tap the microplate to mix the solution evenly;
12. After adding the stop solution 30 Within minutes, measure using a microplate reader. 450nm Set the absorbance value. 540nm or 570nm As a correction wavelength. If dual wavelength correction is not used, the accuracy of the results may be affected;
13. Calculation results: Corrected absorbance values for each standard and sample (OD450-OD540 or OD570) Take the average of the duplicate well readings, and then subtract the average zero standard. OD Value. Four-parameter logic using computer software (4-PL) Curve fitting creates a standard curve. Another method is to plot the standard concentration logarithmically against the corresponding OD Logarithmic curves were generated, and the best fit line was determined by regression analysis. This process produces a sufficiently useful, but not very accurate, data fit. If the sample is diluted, the concentration should be calculated by multiplying by the dilution factor.
Note: The standard curve data provided is for reference only and the sample content should be calculated based on the standard curve drawn for the same test.

III. Kit Parameters
1. Recovery: Different levels of incorporation into cell culture medium samples Mouse IL-2 The recovery rate is determined. The recovery range is 84-107% The average recovery rate is 95% 。
2. Sensitivity: Mouse IL-2 The lowest measurable dose ( MDD ) is generally less than 3.5pg/mL 。 The lowest measurable value is based on 20 The corresponding concentration is calculated by adding the average of the zero-point absorbance values of each standard curve to twice the standard deviation.
3. Correction: This ELISA High-purity recombination of the kit expressed by E. coli Mouse IL-2 Corrected by protein.
4. Linearity: 4 High concentrations were added to different samples. Mouse IL-2 Then use diluent ( 1× ) Dilute the sample within the test range and determine its linearity.

dilution factor average / Expected value ( % ) Scope ( % )
1:2 99 95-105
1:4 100 96-104
1:8 103 97-110
1:16 103 99-109


5. Specificity: This ELISA The method can detect both natural and recombinant substances. Mouse IL-2 Protein. The following factors are mixed with diluent ( 1× ) configured to 50ng/mL Concentration to detect and Mouse IL-2 Cross-reaction. will 50ng/mL The incorporation of interfering factors into the recombination of the intermediate range Mouse IL-2 In the reference substance, to detect the Mouse IL-2 Interference. No significant cross-reactions or interferences were observed.



recombinant mouse protein recombinant human protein
IL-4 IL-2
IL-7 IL-2Rα
 IL-9  
 IL-15  
IL-2Rα  
IL-2Rβ  
IL-2Rγ  
IV. Analysis of Frequently Asked
1. Whiteboard (no color appears after color development is complete)
serial number possible cause Solutions
1 Improper storage of the kit; Mixing of reagents from different kits When purchasing new kits, pay attention to storage conditions; Cannot be mixed
2 Imparts low temperature and short time If the temperature is low, extend the incubation time and color development time.
3 Misaddition or omission of reagents Add the correct reagent strictly according to the instructions.
4 The container used to prepare the solution is not clean, or the water is problematic. Use clean containers and qualified distilled water
5 During the board washing process, the soaking time is long, the board washing times are too many, and the board washing impact is high. Strictly follow the instructions.
6 Non-uniform reagent temperature All reagents must be in equilibrium at room temperature 30 minute
7 detecting antibodies and / or HRP The concentration is too low Refer to the instructions; do not dilute arbitrarily.
2. Flower plate (blank, negative positive control normal, but specimen well OD (The value is significantly high)
serial number possible cause Solutions
1 Few washings, insufficient Wash according to the instructions
2 substrate 3,3',5,5'- Tetramethylbenzidine ( TMB Contamination or exposure to metal ions or oxidizing agents Use clean containers and qualified distilled water during preparation; Protect from light
3 High incubation temperature and / Or too long Control the temperature and time of incubation and final enzymatic reaction
4 The gun tip was not replaced during sample loading, causing cross-contamination. Change the gun tip for each sample.
5 Cross-contamination of nearby holes Vertical patting, use suitable patting paper to avoid paper scraps entering the holes
6 The sample contains endogenous interfering substances. Identify possible infectious substances and handle them accordingly.
7 Sample hemolysis, prolonged storage, incomplete agglutination, bacterial contamination, and additives in the collection tube. Avoid hemolysis, contamination, and prolonged storage.

V. Experimental Flowchart
Species Reactivity Mouse
Theory This kit adopts double antibody sandwich enzyme-linked immunosorbent detection technology. Specific anti-mouse IL-2 antibodies were pre-coated on high affinity enzyme label plates. The standard, the sample to be tested, and the biotinylated detection antibody are added to the well of the enzyme labeled plate. After incubation, the IL-2 present in the sample binds to the solid-phase antibody and the detection antibody to form an immune complex. After washing to remove unbound material, horseradish peroxidase-labeled Streptavidin (Streptavidin-HRP) was added. After washing, a chromogenic substrate was added and the color was developed in the dark. A stop solution was added to stop the reaction, and the absorbance value was measured at a wavelength of 450 nm (reference correction wavelength of 540 nm or 570 nm).
Source Mouse
Synonym aldesleukin, IL2, IL-2, IL-2lymphokine, interleukin 2, interleukin-2, involved in regulation of T-cell clonal expansion, T cell growth factor, T-cell growth factor, TCGF
Wells 0
Composition Please use the kit within its expiration date (new and old products will be shipped randomly).
form Specification Shelf life of diluted or reconstituted reagents after opening
Mouse IL-2 Microplate 1 block After placing the unused slats back into an aluminum foil bag containing a desiccant and sealing them, they can be... 2-8℃ store 30 sky
Mouse IL-2 Standard 2 branch After dissolution, calculate the dosage and repackage, which can be used in -20°C store 14 sky
Mouse IL-2 detecting antibodies 1 branch After the concentrated volume dissolves, it can be... 2-8°C store 14 sky
40×SA-HRP 1 branch 40× The concentration can be 2-8°C storage; 1× Working concentration is not recommended for
10× Buffer for concentration and dilution 1 bottle After opening, it can be 2-8°C store 30 sky
color developing solution 1 bottle
stopping fluid 1 bottle
20× concentrated wash buffer 1 bottle
sealing film 3 Zhang Store at room temperature; do not reuse to avoid contamination.
Background Interleukin 2 (IL-2, also known as TCGF) is a 15-16 kDa monomeric alpha helical glycoprotein belonging to the type I cytokine family. The mouse IL-2 precursor is a 169 amino acid (AA) polypeptide comprising a 20 amino acid signal sequence and a 149 amino acid mature chain. There is also a potential site for O-linked glycosylation, and a polyglutamine region at residues 35-46. It is worth noting that mouse IL-2 has multiple alleles, resulting in differences in the number of glutamines and the number of TSSS repeats at amino acids 21-46 of the mouse IL-2 sequence (SwissProt#: P04351). The amino acid sequence of mature mouse IL-2 is 56% and 73% homologous to that of human-synthesized rat IL-2, respectively, with the main difference being in the polyglutamine region. There is some crossover between mouse and human IL-2. Functionally, IL-2 was originally thought to be a T cell growth factor, and therefore, it is used in the treatment of cancer and AIDS patients, as well as to improve the immune response. However, these efforts have had limited effect. Experimental studies have shown that mice lacking IL-2 or IL-2 receptors can also develop lymphocyte proliferation and autoimmune diseases, indicating that IL-2 is a factor that limits growth rather than induces growth. Recent studies have shown that the real cause of lymphoproliferation is the inability to produce CD4+FoxP3+CD25/IL-2Rα+Treg cells. It is noteworthy that IL-2 is now considered a necessary condition for successful induction of CD8+ memory cell recall responses. It is generally known that mammalian cells expressing IL-2 include CD4+ and CD8+ T cells, NK cells, NKT cells, and dendritic cells. The IL-2 receptor is complex, composed of three different subunits. Two of these subunits can bind to ligands, known as IL-2Rα and IL-2Rβ. IL-2Rα is 55 kDa and binds to IL-2 by virtue of low affinity. IL-2β is 75 kDa, is also a component of the IL-15 receptor, and binds to IL-2 with medium affinity. Signal transduction is accomplished by IL-2Rβ and a common gamma chain (γc) of 64 kDa. γc shares receptors with IL-4, -7, -9, -15, and -21. Signal transduction γc does not bind to IL-2 but heterodimerizes with IL-2Rβ to form a functional IL-2 receptor. The heterotrimeric α-β-γc receptor may bind to IL-2 to form an Rα-Rβ complex.
General Notes 1. Please use the kit within its expiration date.
2. Components of different kits and different batch numbers of kits cannot be mixed.
3. If the sample value is greater than the highest value of the standard curve, the sample should be diluted with diluent (1×) and retested; If the cell culture supernatant sample requires distribution dilution, cell culture medium can be used for other intermediate dilutions except for the final dilution with diluent (1×).
4. Differences in test results can be caused by a variety of factors, including the operation of the experimental personnel, the way the pipette is used, the plate washing technique, the reaction time or temperature, and the storage of the reagent kit.
5. The termination solution in the kit is an acidic solution. Please protect your eyes, hands, face, and clothing when using it.
6. For scientific research only, not for in vitro diagnosis.
Storage Temp. Unopened kit, stored at 2-8°C, shelf life 12 months.
Test Range 15.6pg/mL-1000pg/mL