Human Th1/Th2 & Inflammation 8-Plex Kit (Flow Cytometry Multiplex Bead Assay)

Human Th1/Th2 & Inflammation 8-Plex Kit (Flow Cytometry Multiplex Bead Assay)

Catalog Number: S0Q1003 Reactivity: Human Conjugation: Brand: Starter
Price:
Regular price $0 USD
Regular price Sale price $0 USD
Size:

For shipping services or bulk orders, you may request a quotation.
Secure checkout with
View full details

Product Details

Product Specification


Antigen IL-2、IL-4、IL-5、IL-6、IL-8、IL-10、TNF-a、IFN-γ
Reactivity Human
Stability & Storage

12 months from date of receipt / reconstitution, 2 to 8°C as supplied.

Kit


Precision Intra-assay: <5%;
Inter-assay: <10%
Sample type Serum; Plasma; Cell culture supernatant
Assay type Sandwich (quantitative)
Sensitivity <2.0 pg/mL
Recovery 70% - 130%
Assay time 70 minutes
Species reactivity Human

Background

The Cytometric Bead Array (CBA) technology enables the capture of soluble analytes or analyte panels using beads with known sizes and fluorescence intensities, thereby allowing the detection of analytes via flow cytometry.

Each capture bead in the CBA kit is conjugated with a specific antibody. The detection reagent provided with the kit is Streptavidin-Phycoerythrin, and the intensity of the fluorescent signal generated by this reagent is proportional to the amount of bound analyte.

When the capture beads and detection reagent are incubated together with the test sample containing target analytes, sandwich complexes are formed. Detection of these complexes by flow cytometry allows for the identification of particles exhibiting the fluorescence characteristics of both the beads and the detection reagent.

Picture

CBA

Distribution Diagram
In the figure, the R1 gate represents the selected target capture microsphere group, which is used to define this group for subsequent detection and analysis of fluorescence signals.

Distribution Diagram
The figure shows the distribution of microsphere clusters corresponding to different cytokines; it enables the simultaneous differentiation of multiple cytokines, providing a basis for the group definition for the subsequent quantitative analysis of each cytokine.

Distribution Diagram
The histogram reflects the number of particles in each microsphere group.

Distribution Diagram
This figure shows a typical flow cytometry scatter plot of the CBA quantitative detection reagent. The captured microspheres with different APC fluorescence intensities are sequentially enclosed (A2 to A9), corresponding to the 8 different target substances detectable in the kit; the signal intensity of the PE channel is proportional to the content of the target substances bound to each microsphere.

Standard curve
Example of IL-2,IL-4,IL-5,IL-6,IL-8,IL-10,TNF-a,IFN-γ standard curve in Assay Diluent .

Linearity
After serial dilution of the high-concentration combined proteins of human IL-2, IL-4, IL-5, IL-6, IL-8,IL-10, TNF-a,IFN-γ using plasma, serum, and cell culture supernatant matrices respectively, the concentrations of human IL-2, IL-4, IL-5, IL-6, IL-8,IL-10, TNF-a,IFN-γ were measured and interpolated from the target standard curves and corrected for sample dilution.

Intra-assay precision
Ten replicates of each of three different levels of IL-2, IL-4, IL-5, IL-6, IL-8,IL-10, TNF-a and IFN-γ were tested.

Inter-assay precision
Ten repeated tests were conducted on three different levels of IL-2, IL-4, IL-5, IL-6, IL-8, IL-10,TNF-a and IFN-γ using different batches of reagents.

Specificity
The antibodies used in the kit have been screened for their specific reactivity with particular cytokines. After analyzing samples containing only a single recombinant cytokine protein, it was found that when using this detection method, there was no cross-reaction or background detection of cytokines in other capture magnetic bead groups.

Theoretical limit of detection
Repeat the detection of the blank sample 20 times using the same batch of reagents, and calculate the theoretical limit of detection.

Recovery
Within the detection range, the combined cytokine proteins were added to different matrices at medium and low concentrations.
The matrices used in these experiments were diluted 2-fold with the assay buffer before the addition of the cytokine proteins.
The results were compared with those of samples that had the same concentration of cytokines added in the assay buffer.

Protocol Diagram