Standard Curve
Example of Human MMP-9 standard curve in Assay Diluent A1. Plotted are the background-subtracted data.
This standard curve is provided for demonstration only.
Product Details
Product Details
Product Specification
| Antigen | MMP-9 |
| Immunogen | Recombinant Protein |
| Antibody Type | Recombinant mAb |
| Reactivity | Hu |
| Purification | Protein A |
| Stability & Storage | 12 months from date of receipt / reconstitution, 2 to 8°C as supplied. |
Kit
| Precision | Intra-assay: 2.9%; Inter-assay: 5.3% |
| Sample type | Serum; EDTA Plasma, Heparin Plasma, Citrate Plasma |
| Assay type | Sandwich (quantitative) |
| Sensitivity | 9.37 pg/mL |
| Range | 46.9 pg/mL – 3,000 pg/mL |
| Recovery | Serum: 99% EDTA Plasma: 102% Citrate plasma: 97% Heparin Plasma: 90% |
| Assay time | 60 minutes |
| Species reactivity | Hu |
Background
Human matrix metalloproteinase-9 (MMP-9), also known as gelatinase B or 92 kDa type IV collagenase, is a zinc-dependent endopeptidase encoded by the MMP9 gene that is secreted as an inactive zymogen and activated upon propeptide cleavage, and which—though largely absent from healthy tissues at baseline—is robustly induced during inflammation, tissue remodeling, and cancer to degrade gelatins, type IV and V collagens, and fibronectin, thereby breaking down basement membrane barriers and regulating leukocyte migration, angiogenesis, and inflammatory signaling; elevated MMP-9 levels are clinically associated with tumor progression and metastasis, cardiovascular remodeling, neuroinflammation with blood-brain barrier disruption, and inflammatory bowel disease, making it an active target for selective inhibitor development.
Picture
Picture
ELISA
Spike Recovery
Recovery of Rat IgE was assessed by spiking a 3000 pg/mL standard into Human serum (0.2%), Human plasma-EDTA (0.2%), Human plasma-Citrate (0.2%) and Human plasma-Heparin (0.2%). Serial dilutions of the spiked sample were prepared, and the percent recovery was calculated for each dilution relative to the expected concentration.
Dilution Linearity
The concentrations of Human MMP-9 were measured and interpolated from the target standard curves and corrected for sample dilution.
The leftmost bar in the graph shows biological samples diluted 500-fold in Assay Diluent. "Undiluted" was defined as biological sample diluted 500-fold. The mean target concentration was determined to be 281.1 ng/mL in Human serum; The mean target concentration was determined to be 35.09 ng/mL in Human plasma-EDTA; The mean target concentration was determined to be 184.07 ng/mL in Human plasma-Citrate; The mean target concentration was determined to be 40.70 ng/mL in Human plasma-Heparin.
Intra-Assay Precision (Precision within an assay)
Three samples of known concentration were tested fifteen times on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays).
Three samples of known concentration were tested in separate assays to assess inter-assay precision. Assays were performed with at least three lots of components.
Determination of Minimum Detectable Dose (MDD)
The MDD was determined using three independent lots of assay components. For each lot, 19 replicate measurements of the diluent (zero calibrator) were performed. The mean (AVERAGE) and standard deviation (STDEV) of the 19 replicates were calculated. The MDD for each lot was then calculated according to the following formula:
MDD = 2 × STDEV + AVERAGE
Cross‑reactivity
The heatmap displays OD450 ratios for serum samples from multiple species. Color intensity is proportional to the O.D.450 ratio (test species / reference species). Darker colors indicate higher cross‑reactivity.
HOOK Effect Threshold
The upper limit of the HOOK effect was established at 100× the highest calibrator concentration (equivalent to a 2-log10 increase). No HOOK effect was observed below this threshold, confirming that samples within this concentration range yield reliable quantitative results without signal depression.
Protocol Diagram