| Usage |
1. Users need to bring their own materials; reagent kits are not provided. Able to measure absorbance 450nm microplate reader Enzyme plate thermostatic shaker or other horizontal shaker Multi-channel and single-channel pipettes deionized water centrifuge tube or 96 pore dilution transfer plate
2. Reagent preparation Allow all reagents to return to room temperature before use. This kit comprises a 96 The amount of reagent required for the well plate.
1X Wash Buffer: Dilute with deionized water 10X Wash Buffer , preparation 1X Wash Buffer 。 will 10mL 10X Wash Buffer with 90mL Made by mixing deionized water 100mL 1X Wash Buffer Gently stir until smooth. Note: 10X Wash Buffer In 2-8ºC Crystals may precipitate during storage, which is a normal phenomenon and will completely dissolve after returning to room temperature. Please wait until the crystals are completely dissolved before diluting with deionized water.
1X Detector Antibody: use Assay Diluent #12 dilution 100X Detector Antibody , preparation 1X Detector Antibody 。 will 100 μL 100X Detector Antibody with 9900 μL Assay Diluent #12 Mix evenly, that is, prepare into 10 mL 1X Detector Antibody Working fluid. The antibody working solution is prepared and used immediately. It is recommended to use the working solution on the same day after it is prepared, and it needs to be reprepared the next day.
standard preparation A new set of standards was prepared for each experiment. The lyophilized standard powder becomes the standard mother liquor after dissolution. If stored for a long time, it needs to be repackaged promptly. -80℃ Save. The standard working solution is prepared and used immediately. It is recommended to use the working solution within one hour after preparation. After one hour, it needs to be reprepared. The preparation process of the recommended standard curve for this kit is described below.
IMPORTANT : The identified mass and redissolution volume of the standard powder in each tube are indicated on the standard label. Before opening the cap, centrifuge briefly before adding the volume indicated on the tube label. Assay Diluent #12 Dissolution is Human IL-10 Standard mother liquor. Fully dissolved at room temperature 10 Minutes and mix gently. This is 5000 pg/mL Standard mother liquor. 1 Mark with a marker 8 One column of centrifuge tubes or dilution plates ( 8 (holes), respectively standard working solution 1– 8 。 2 ,in 1 Number pipe / Add to the hole 400 μL Assay Diluent #12 , in 2-8 Number pipe / Add to the hole 250 μL Assay Diluent #12 。 3 Prepare a standard gradient using the standard stock solution as shown in the figure below. Standard #8 For blank control ;
3. Sample Preparation serum : Samples should be collected into serum separation tubes. After the clot forms, 2000 x g centrifuge 10 Minutes, serum samples were collected. The serum samples were used... Assay Diluent #12 After dilution, the sample was loaded. Please provide undiluted serum samples at -20°C Or store at a lower temperature 3 months. Avoid repeated freeze-thaw cycles. Cell or tissue culture supernatant : The cell or tissue culture supernatant is placed in 2000 x g Lower centrifugation 10 Minutes, remove debris. After collecting the supernatant, it can be applied directly or used Assay Diluent #12 After dilution, the sample was loaded. In -20°C Or preserve undiluted samples below. Avoid repeated freeze-thaw cycles. Enzyme label plate preparation : 96 The well removable enzyme label plate is ready-to-use and can be loaded directly without further processing. The plate shape is 12 strip 1 x 8 slats. Unused slats should be immediately returned to the foil bag containing the desiccant packaging, resealed, and stored. 2-8°C Save now.
IV. Experimental All materials and reagents will need to be equilibrated to room temperature prior to use. It is recommended that all standards, controls and samples be tested in duplicate in parallel wells. 1 Prepare all reagents, standard working solutions, and samples according to the instructions on the previous page. 2 Remove excess enzyme-labeled strips from the removable plate frame, place them back in the foil bag containing the desiccant packaging, reseal, and return. 2-8℃ Storage. 3 , take 100µL The sample or standard is added to the corresponding well. 4 Apply the sealing film, under the following setting conditions: “ rotation speed 700 turn / Minutes and temperature 25ºC” Incubation on an enzyme-labeled plate constant temperature shaker 60 minutes. 5 ,use 300µL 1X Wash Buffer Wash each well three times. After the last wash, turn the plate upside down and pat it gently several times on a clean paper towel to remove excess liquid. 6 Add to each well 100µL 1XDetector Antibody Working fluid. 7 Apply the sealing film, under the following setting conditions: “ rotation speed 700 turn / Minutes and temperature 25ºC” Incubation on an enzyme-labeled plate constant temperature shaker 60 minutes. 8 After incubation, use 300µL 1X Wash Buffer Wash each well three times. After the last wash, pat gently several times on a clean paper towel to remove excess liquid. 9 Add to each well 100µL TMB substrate, in 500 turn / The enzyme label plate was developed in a constant temperature shaker or shaker for approximately 10 minutes, protected from light. 15 minutes. Given the variability of laboratory environmental conditions, the optimal development time may be... 5 to 15 Changes between minutes. Note : After adding the stop solution, the color changed from blue to yellow, and the signal intensity increased by approximately 3 Double. To avoid signal saturation, the enzyme label plate can be pre-read before adding the stop solution to achieve blue color at high standard concentrations. O.D.650 or O.D.620 = 1.0 Before proceeding to the next step. 10 Add to each well 100µL Stop solution, mix well and then... 30 Measure each well within minutes OD450 Use the settings as 450 nm The microplate reader. If there is wavelength correction, set it to 540nm,570nm or 620nm , after calibration OD The value is 450 nm Subtract the measured value of the correction wavelength from the measured value.
V. Result Calculation Take the average of the replicate readings for each standard and sample and subtract the average reading of the blank control. Four-parameter logic can be generated using (4-PL) Curve-fitting computer software creates standard curves and scales sample concentrations. If the sample is diluted before loading, the concentration calculated from the standard curve must be multiplied by the dilution factor to give the actual concentration of the sample.
Example of standard curve formula 
enzyme label board layout You can use this diagram to layout and record standards and analytical samples VI. Kit Parameters Precision: Intra-batch precision: 4.1% ; Inter-batch precision: 5.4% Recovery: Cell culture supernatant: 107% ; Serum: 108% Detection time: 2.5 hour
|