Linearity
After serial dilution of the high-concentration combined proteins of IL-1β,IL-2, IL-4, IL-5, IL-6, IL-8,IL-10, IL-13,TNF-α,and IFN-γusing plasma, serum, and cell supernatant matrices respectively, the linear slopes between the actual detected values and the theoretical values are as shown in the table.
Product Details
Product Details
Product Specification
| Antigen | IL-1β、IL-2、IL-4、IL-5、IL-6、IL-8、IL-10、IL-13、TNF-α、IFN-γ |
| Antibody Type | Recombinant mAb |
| Reactivity | Human |
| Stability & Storage | 12 months from date of receipt / reconstitution, 2 to 8℃ as supplied. |
Kit
| Precision | 批内变异系数 (Intra-assay): 2.67%-7.76%; 批间变异系数 (Inter-assay): 8.3%-13.76% |
| Sample type | Cell culture supernatant, Serum, Plasma |
| Assay type | Sandwich (Qualitative) |
| Sensitivity | <1.0pg/ml |
| Range | 0.5-10000 pg/mL |
| Recovery | 70%-130% |
| Assay time | 70 minutes |
| Species reactivity | Human |
| Plate | Non-detachable plate |
Background
xMAP Technology uses colored beads to perform biological assays similar to ELISA or nucleic acid hybridization assays. By color-coding microscopic beads into many spectrally distinct sets, each bead set can be coated with a nucleic acid or protein capture molecule specific to a particular biological target, allowing the simultaneous capture of multiple analytes from a single sample. Because of the microscopic size and low density of these beads, assay reactions exhibit virtually solution-phase kinetics. However, once an assay is complete the solid phase characteristics allow each bead to be analyzed discretely. By incorporating magnetic properties into xMAP microspheres, assay washing is simplified while maintaining desirable solution-phase properties.
Picture
Picture
Luminex

Standard Curve
Example of IL-1β, IL-2,IL-4,IL-5,IL-6,IL-8,IL-10,IL-13,TNF-α,IFN-γstandard curve in Assay Diluent.
Performance Characteristics
Precision
Ten replicates of each of three different levels of IL-1β, IL-2,IL-4,IL-5,IL-6,IL-8,IL-10,IL-13,TNF-α,IFN-γwere tested.
Specificity
The antibodies used in the kit have been screened for their specific reactivity with particular cytokines. After analyzing samples containing only a single recombinant cytokine protein, it was found that when using this detection method, there was no cross-reaction or background detection of cytokines in other capture magnetic bead groups.

Recovery
Within the detection range, the combined cytokine proteins were added to serum,plasma,cell supernatant at high concentrations.The matrices used in these experiments were diluted 2-fold with the assay buffer before the addition of the cytokine proteins. Theresults were compared with those of samples that had the same concentration of cytokines added in the assay buffer.
Distribution Diagram(Luminex)
Luminex Distribution diagram of Beads on the instrument operation software interface.
Distribution Diagram(CBA)
Distribution diagram of Beads on the interface of flow cytometry operating software.
Distribution Diagram(CBA)
Grouping diagram of Beads on the interface of flow cytometry operating software.
Distribution Diagram(CBA)
Grouping diagram of Beads on the interface of flow cytometry operating software.