Human IL-2R Kit (HICA)

Human IL-2R Kit (HICA)

Catalog Number: UA086036 Brand: UA BIOSCIENCE
Price:
Regular price $1,599 USD
Regular price Sale price $1,599 USD
Size:

For shipping services or bulk orders, you may request a quotation.
Secure checkout with
View full details

Product Details

Product Specification


Stability & Storage

Store at 2~8°C protected from light for 18 months; after reconstitution, the standard solution can be aliquoted and stored at -20°C, avoiding repeated freeze-thaw cycles.

Background

Principle of the Assay:

This kit employs a homogeneous immuno chemiluminescence assay (HICA) based on the dual-antibody sandwich principle for the detection of cytokine concentrations. The operation is simple and requires no washing steps.

The detection system consists of two types of microspheres: receptor microspheres conjugated with antibody 1 targeting the protein of interest, and donor microspheres conjugated with streptavidin, along with biotin-labeled antibody 2. During the reaction, the target protein binds with the antibodies and microspheres to form an immune complex, bringing the two types of microspheres into close proximity. When the distance between the donor and receptor microspheres is less than 200 nm, singlet oxygen generated by light excitation can transfer to the receptor microspheres and trigger chemiluminescence. Conversely, if the target protein is absent, the distance between the microspheres is too large, and no signal is produced.

By measuring the intensity of the chemiluminescent signal, quantitative analysis of the target protein can be achieved. This method is characterized by its simplicity, rapid reaction, and high sensitivity.

Components

Name

Specifications

Component Specifications

Test ReagentR1A

500T

1mL/per bottle ×1

2000T

4ml/per bottle ×1

10000T

20ml/per bottle ×1

Test reagentR1F

500T

1mL/per bottle ×1

2000T

4ml/per bottle ×1

10000T

20ml/per bottle ×1

Test reagentR2

500T

1mL/per bottle ×1

2000T

4ml/per bottle ×1

10000T

20ml/per bottle ×1

Test reagentR3

500T

6mL/per bottle ×1

2000T

24ml/per bottle ×1

10000T

120ml/per bottle ×1

Standard

500T

0.05μgLyophilized product ×1

2000T

0.05μgLyophilized product ×2

10,000T

0.05μgLyophilized product ×5

Reference standardBuffer solution

500T

6mL pervial ×1

2000T

12ml pervial ×1

10,000T

30ml perbottle ×1


Note: It is recommended to use microplates for the assay (384or96well plates, white, with shallow wells)

Protocol

I. Sample Requirements
1. To remove impurities from the sample, samples must be centrifuged prior to detection (1000 ×g, 10 minutes).
2. When detecting target proteins in serum samples, the Standard should be diluted using negative serum with a background value below the detection limit, and a standard curve should be established accordingly to calculate the actual concentration.
3. If the measured concentration of a sample exceeds the highest value on the standard curve, it must be appropriately diluted and retested to ensure the result falls within the valid measurement range.

II. Detection Protocol

2.1 Standard Gradient Sample Preparation: Reconstitute the lyophilized Standard with 50μL of deionized water, then dilute using Standard Buffer. The recommended dilution scheme is shown in the table below:

Gradient

Concentration (pg/mL)

Standard(μL)

Diluent (μL)

C12

100000

20μL Calibrator

180

C11

30000

60μL C12

140

C10

10000

60μL C11

120

C9

3000

60μL C10

140

C8

1000

60μL C9

120

C7

300

60μL C8

140

C6

100

60μL C7

120

C5

30

60μL C6

140

C4

10

60μL C5

120

C3

3

60μL C4

140

C2

1

60μL C3

120

C1

0

—

120


 

 

2.2 Detection Protocol: Protocols 1 and 2 are applicable to most assays. Protocol 3 is suitable for the detection of samples containing Biotin, such as RPMI medium.

Detection Protocol

Detection Protocol 1 (37℃ Rapid Detection)

Detection Protocol 2 (Room Temperature Detection)

Detection Protocol 3 (Suitable for Matrices Containing Biotin)

​Step 1:

Take 2µL of sample, add 6µL of pre-mixed R1A+R1F+R2*

Take 2µL of sample, add 6µL of pre-mixed R1A+R1F+R2*

First, mix R1F, R2, and R3 at a ratio of 1:1:1, and let stand for 30 minutes; take 3µL of sample, add 13µL of pre-mixed reagent, protect from light/green light

Incubation

​Shake/mix at 400 rpm for 1 min, incubate at 37℃ for 15 minutes

​Shake/mix at 400 rpm for 1 min, ​incubate at room temperature for 60 minutes

Shake/mix at 400 rpm for 1 min, incubate at 37℃ for 15 minutes, protect from light/green light

​Step 2:

Add 12µL R3, protect from light/green light

Add 12µL R3, protect from light/green light

Add 3µLR1A, protect from light/green light

Incubation

​Shake/mix at 400 rpm for 1 min, incubate at 37℃ for 10 minutes, protect from light/green light

​Shake/mix at 400 rpm for 1 min, ​incubate at room temperature for 30 minutes, protect from light/green light

​Shake/mix at 400 rpm for 1 min, ​incubate at 37℃ for 10minutes, protect from light/green light

Reading​

Read on instrument, protect from light/green light

Read on instrument, protect from light/green light

Read on instrument, protect from light/green light


 

* Note: Prior to testing, pre-mix R1A, R1F and R2 at a volume ratio of 1:1:1, and use the mixture within 1 hour after mixing.

* If testing RPMI samples, you may additionally add RPMI enhancer solution (Catalog No.: UA086145). Pre-mix 50 μL of R3 with 100 μL of RPMI enhancer solution before adding, which significantly improves reagent performance.


III.Performance Testing

3.1 Example of Complete Standard Curve:

 

3.2 Validation of Performance Parameters:

LOD: Detect the Standard C1 repeatedly for 20 times, calculate the mean signal and SD, use the standard curve to calculate the concentration corresponding to mean signal + 2×SD, which is the Limit of Detection (LOD).

Detection Protocol

Matrix

LOD(pg/mL)

Quantitative Range(pg/mL)

Dynamic Range(pg/mL)

Detection Protocol 1

Buffer

2.95

3~30000

0~100000

DMEM

0.12

0.1~30000

Fetal Bovine Serum

7.21

7.2~30000

Detection Protocol2

Buffer

15.01

15~30000

Detection Protocol3

RPMI

21.79

21.8~30000


 


Precision (Precision):
Intra-assay Precision (Intra-Precision): On the same plate within a single experiment, samples at known low, medium, and high concentrations are each detected repeatedly for 10 times to evaluate the intra-assay precision of this assay. Both standards and samples are tested using different protocols. The coefficient of variation (%CV) for all concentration levels is below 10%, indicating that this method has good reproducibility within the same batch of tests.

Detection Protocol

Matrix

Precision

Low Concentration

High Concentration

Detection Protocol 1

Buffer

3.5%

3.1%

DMEM

2.5%

4.8%

Fetal Bovine Serum

2.1%

2.0%

Detection Protocol2

Buffer

6.0%

1.7%

Detection Protocol3

RPMI

5.0%

4.4%


 


Inter-assay Precision (Inter-Precision): Samples at known low, medium, and high concentrations are detected in 5 independent experiments. Standards and samples are both tested using Detection Protocol 1 with 5 replicates. The results for all concentration levels show a low coefficient of variation (%CV < 10%), indicating that this method has good reproducibility across different batches of tests.

Detection Protocol

Matrix

Inter-assayPrecision

Low Concentration

High Concentration

Detection Protocol 1

Buffer

10.0%

6.6%


 

Accuracy (Recovery):

The recovery rate was tested after mixing high-concentration and low-concentration quality controls at a ratio of 1:9. All values fall within 80%-120%, indicating that this method has good accuracy.

Detection Protocol

Matrix

Recovery Rate

Detection Protocol 1

Buffer

86.6%

DMEM

85.4%

Fetal Bovine Serum

87.6%

Detection Protocol2

Buffer

86.3%

Detection Protocol3

RPMI

86.8%


 

Specificity: The following proteins were diluted to 30 n g/mL using Standard Buffer to test cross-reactivity rates.

Detection Protocol

Analyte

Cross-reactivity Rate

Detection Protocol 1

Human IL-2

0.00%

Human IL-5

0.00%

Human IL-8

0.00%

Human IL-10

0.10%


 

Traceability: Tested against reference material NIBSC 97/600. 1 IU/mL ≈ 68 pg/mL.

Guidelines



Reagent R3 must be protected from light. Sample addition and incubation should be performed under green light (<100 LUX).

Recalibration is required for each test. Each concentration point of the standard should be tested in at least duplicate, and a four-parameter (weighting 1/Y²) curve fitting should be applied for calculation.

Temperature and time must be carefully controlled during incubation. Microplates should be covered with a sealing film, and a microplate reader with ALPHA function is recommended.

The dilution matrix for calibrators should be consistent with the sample matrix. Reconstituted calibrators must be used within 2 hours.

Components from different reagent kit lots must not be mixed.