Product Details

Product Specification


Stability & Storage

Store at 2~8°C protected from light for 18 months; after reconstitution, the standard solution can be aliquoted and stored at -20°C, avoiding repeated freeze-thaw cycles.

Background

Test Principle:

This kit employs a homogeneous immuno chemiluminescence assay (HICA) based on the double-antibody sandwich method for the detection of cytokine concentrations. The operation is simple and requires no washing steps.

The detection system includes two types of microspheres: acceptor microspheres conjugated with antibody 1 targeting the protein of interest, and donor microspheres conjugated with streptavidin, along with biotin-labeled antibody 2. During the reaction, the target protein binds with the antibodies and microspheres to form an immune complex, bringing the two microspheres into close proximity. When the distance between the donor and acceptor microspheres is less than 200 nm, singlet oxygen generated upon light excitation can transfer to the acceptor microsphere and trigger chemiluminescence. Conversely, if the target protein is absent, the distance between the microspheres is too large, resulting in no signal.

By measuring the intensity of the chemiluminescent signal, quantitative analysis of the target protein can be achieved. This method offers advantages such as simplicity, rapid reaction, and high sensitivity.

Components

Name

Specification

Component Specification

hIL-2 Assay Reagent R1

100 Tests

400 μL/bottle × 1

500 Tests

2 mL/bottle × 1

2,000 Tests

8 mL/bottle × 1

10,000 tests

40 mL/bottle × 1

hIL-2 assay reagent R2

100 tests

400 μL/bottle × 1

500 tests

2 mL/bottle × 1

2,000 tests

8 mL/bottle × 1

10,000 tests

40 ml/bottle × 1

hIL-2 assay reagent R3

100 tests

1 mL/bottle × 1

500 tests

5 mL/bottle × 1

2,000 tests

20 mL/bottle × 1

10,000 tests

100 mL/bottle × 1

hIL-2 standard

100 tests

1.67 μg lyophilized powder × 1

500T

1.67μg lyophilized product × 1

2000T

1.67μg lyophilized product × 2

10000T

1.67μg lyophilized product × 5

hIL-2 standard buffer

100T

1.2mL/bottle × 1

500T

6 mL/bottle × 1

2000T

12 mL/bottle × 1

10,000T

30 mL/bottle × 1


 

Note: It is recommended to use microplates (384- or 96-well plates, white, with shallow wells) for this assay.

Protocol

I. Sample Requirements
1. To remove impurities from the sample, centrifuge the sample before testing (1000 ×g, 10 minutes).
2. When detecting target proteins in serum samples, the standard should be diluted using negative serum with a background value below the lower limit of detection, and a standard curve should be established accordingly to calculate the actual concentration.
3. If the measured concentration of a sample exceeds the highest value of the standard curve, it must be appropriately diluted and retested to ensure the results fall within the valid measurement range.

II. Detection Procedure

2.1 Preparation of Standard Gradient Samples Reconstitute the lyophilized standard with 50μL of deionized water, then dilute using the standard buffer. The recommended dilution scheme is shown in the table below (adjustable according to specific conditions):

Gradient

Concentration (pg/mL)

Standard (μL)

Diluent(μL)

C12

100000

10μL Standard

324

C11

30000

60μL C12

140

C10

10000

60μL C11

120

C9

3000

60μL C10

140

C8

1000

60μL C9

120

C7

300

60μL C8

140

C6

100

60μL C7

120

C5

30

60μL C6

140

C4

10

60μL C5

120

C3

3

60μL C4

140

C2

1

60μL C3

120

C1

0

—

120


 

 

2.2 Detection Procedure: Procedure 1 and Procedure 2 are suitable for the vast majority of tests, while Procedure 3 is suitable for samples containing Biotin, such as RPMI medium.

Detection Procedure

Detection Procedure1 (37°C Rapid Detection)

Detection Procedure2 (Room Temperature Detection)

Detection Procedure3 (Suitable for Biotin-containing Matrices)

Step 1:

Add 2µL sample to 8µL pre-mixed R1+R2*

Add 2µL sample to 8µL pre-mixed R1+R2*

First, mix R1, R2, and R3 at a 1:1:1 ratio, let stand for 30 minutes, protect from light/green light

Incubation

Mix by shaking at 400 rpm for 1 min, incubate at 37°C for 15 minutes

Mix by shaking at 400 rpm for 1 min, incubate at room temperature for 60 minutes

 

Step 2:

Add 10µL R3, protect from light/green light

Add 10µL R3, protect from light/green light

Add 2µL sample to the above 8µL pre-mixed reagents, protect from light/green light

Incubation

Mix by shaking at 400 rpm for 1 min, incubate at 37°C for 10 minutes, protect from light/green light

Mix by shaking at 400 rpm for 1 min, incubate at room temperature for 30 minutes, protect from light/green light

Mix by shaking at 400 rpm for 1 min, incubate at 37°C for 25 minutes, protect from light/green light

Readout

Read on instrument, protect from light/green light

Read on instrument, protect from light/green light

Read on instrument, protect from light/green light


 

* Note: Before testing, pre-mix R1 and R2 at a volume ratio of 1:1. Use within 1 hour after mixing.

* If testing RPMI samples, you may additionally add RPMI enhancer solution (Catalog No.: UA086145). Pre-mix 50 μL of R3 with 100 μL of RPMI enhancer solution before adding, which significantly improves reagent performance.


III. Performance Testing

3.1 Example of Complete Standard Curve:

 

 


3.2 Performance Parameter Validation:

LOD: Repeat the detection of Standard C1 20 times, calculate the mean signal and SD, and use the standard curve to calculate the concentration corresponding to Mean Signal+2×SD, which is the Limit of Detection (LOD).

Detection Procedure

Matrix

Limit of Blank(pg/mL)

Quantitative Range(pg/mL)

Dynamic Range(pg/mL)

Procedure 1

Buffer

3.26

3.3~30000

0~100000

DMEM

15.28

15.3~30000

Fetal Bovine Serum

0.66

0.6~30000

Procedure 2

Buffer

9.08

9.1~30000

Procedure 3

RPMI

62.83

62.8~30000


 


Precision (Precision):
Intra-Precision: On the same plate in a single experiment, samples with known low, medium, and high concentrations are each tested 10 times to evaluate the intra-assay precision. Standards and samples are tested using different procedures. The coefficient of variation (%CV) for all concentration samples is below 10%, indicating good reproducibility of this method within the same batch.

Detection Procedure

Matrix

Repeatability

Low Concentration

High Concentration

Procedure 1

Buffer

2.9%

2.4%

DMEM

2.6%

1.1%

Fetal Bovine Serum

3.6%

1.2%

Procedure 2

Buffer

3.6%

1.1%

Procedure 3

RPMI

9.6%

3.4%


 


Inter-Precision: Samples with known low, medium, and high concentrations are tested in 5 independent experiments. Standards and samples are both tested using Procedure 1 with 5 replicates. The test results for all concentration samples show a low coefficient of variation (%CV<10%), indicating good reproducibility of this method across different batches.

Detection Procedure

Matrix

Inter-assay Precision

Low Concentration

High Concentration

Procedure 1

Buffer

3.3%

2.4%


 

Accuracy (Recovery):

High-concentration and low-concentration quality control samples are mixed at a 1:9 ratio and tested for recovery rate. The recovery rates are all between 80%-120%, indicating good accuracy of this method.

Detection Procedure

Matrix

Recovery Rate

Procedure 1

Buffer

88.1%

DMEM

88.5%

Fetal Bovine Serum

88.3%

Procedure 2

Buffer

96.0%

Procedure 3

RPMI

89.7%


 

Specificity: The following proteins are diluted to 3μg/mL using standard buffer to test the cross-reactivity rate.

Detection Procedure

Test Substance

Cross-Reactivity Rate

Procedure 1

Mouse IL-2

0.00%


 

Traceability: The reference standard NIBSC 86/500 was tested, where 1IU/mL≈105pg/mL.

Guidelines



Reagent R3 must be protected from light. Sampling and incubation should be performed under green light (<100 LUX).

Recalibration is required for each test, with at least duplicate wells for each standard concentration point. A four-parameter (weighting: 1/Y²) fitting calculation should be applied.

Temperature and time must be controlled during incubation. Microplates should be covered with a film, and a microplate reader with ALPHA function is recommended.

The dilution matrix for calibrators should match the test samples. Reconstituted calibrators must be used within 2 hours.

Components from different reagent kit lots must not be mixed.