Product Details
Product Details
Product Specification
| Usage |
Specimen Requirements
|
||||||||||||||||||||||||||||||||||||
| Sensitivity | 0.05ng/ml | ||||||||||||||||||||||||||||||||||||
| Species Reactivity | Human | ||||||||||||||||||||||||||||||||||||
| Theory | This kit uses a double-antibody sandwich assay to measure the level of human platelet membrane glycoprotein Ib/IX complex (GPIb/IX) in a specimen. A microplate is coated with purified human platelet membrane glycoprotein Ib/IX complex (GPIb/IX) antibody to prepare a solid-phase antibody. Platelet membrane glycoprotein Ib/IX complex (GPIb/IX) is then added sequentially to the antibody-coated microwells. The antibody then binds to HRP-labeled platelet membrane glycoprotein Ib/IX complex (GPIb/IX) antibody, forming an antibody-antigen-enzyme-labeled antibody complex. After thorough washing, the substrate TMB is added for color development. TMB converts to blue under the catalysis of HRP and to yellow under the action of acid. The intensity of the color is positively correlated with the level of platelet membrane glycoprotein Ib/IX complex (GPIb/IX) in the sample. The absorbance (OD value) was measured at a wavelength of 450 nm using a microplate reader, and the concentration of human platelet membrane glycoprotein Ib/IX complex (GPIb/IX) in the sample was calculated using the standard curve. | ||||||||||||||||||||||||||||||||||||
| Detection Type | This kit is used to determine the content of platelet membrane glycoprotein Ib/IX complex (GPIb/IX) in human serum, plasma and related liquid samples. | ||||||||||||||||||||||||||||||||||||
| Composition |
|
||||||||||||||||||||||||||||||||||||
| General Notes | 1. After removing the kit from the refrigerator, allow it to equilibrate at room temperature for 1 hour before use. If the enzyme-coated plate is opened and not completely used, store it in a sealed bag. 2. Crystals may form in the concentrated wash buffer. Warming in a water bath can aid dissolution during dilution. This will not affect the results. 3. Use a pipette for each sample addition step and frequently calibrate its accuracy to avoid experimental error. The time for each addition should ideally be within 5 minutes. For large numbers of samples, using a dispenser is recommended. 4. Develop a standard curve with each measurement, preferably with replicates. If the analyte content in the sample is too high (the sample OD value is greater than the OD value of the first standard well), dilute the sample a certain number of times (n times) with sample diluent before measurement. When calculating the final dilution, multiply by the total dilution factor (×n×5). 5. The sealing film is for single use only to avoid cross contamination. 6. Please store the substrate in a dark place. 7. Strictly follow the instructions for operation. The test results must be determined based on the readings of the microplate reader. 8. All samples, washing solutions and various wastes should be treated as infectious materials. 9. Components of different batches of this reagent must not be mixed. |
||||||||||||||||||||||||||||||||||||
| Storage Temp. | If the unopened test kit is sealed and stored at 2-8°C, it is valid for 6 months. | ||||||||||||||||||||||||||||||||||||
| Test Range | 0.2ng/ml - 9ng/ml |